Installation
When installing the board, perform the following steps:
a.Remove the top and inner cover of the power supply as discussed in Section 3 under Repair and Replacement.
b.Remove the plate next to the barrier strip on the rear panel of the supply by unscrewing the 2 M3 screws.
c.Insert the already prepared 002 board in the slot closest to the right side (looking from the
d.Use the two M3 screws to connect the rear end of the 002 board to the rear panel of the supply.
e.Attach ribbon cables from the A2 Control Board A2J1 to A7J1 and A2J2 to A7J2.
f.Replace the inner and outer cover of the supply.
Connector Assembly Procedure
The following instructions describe assembly of the mating connector provided to interface the user's system with the option connector, J3. Figure
Proceed as follows:
Note: It may be desirable to set up a test interface before final assembly of the mating connector to allow checkout of the system. A mating connector with pins accessible for temporary wiring is available from Agilent Technologies, Agilent part number
If the cable assembly presents RFI or ESD problems, a shielded cable assembly accessory Agilent Part Number
a.If a
b.Strip 3/16 inch of insulation from the end of each wire to be used.
c.Insert each wire into a contact pin (1) and crimp firmly.
d.Insert each pin into a proper hole in
Note | Once the pins are locked into the |
| Therefore, be certain pin is in proper hole before inserting fully. |
e.Screw a slotted setscrew (3) partially into a square nut (4) and place in position in connector shield assembly (6).
f.Place strain relief (5) in position in connector shield assembly (6), just under set screw (3). Be certain that strain relief is oriented as shown in Figure
g.Place connector pin housing (2) in shield assembly (6) and route cable through cable entrance.
h.Fold connector assembly (6) and secure with three screws.
i.Strain relief set screw (3) can now be adjusted from top of connector to clamp firmly on cable.
j.Clip fasteners (7) onto ends of connector pin housing (2).
k.Connector can now be plugged onto option connector J3 and secured with two screws (8) into the threaded
83