0.25Ω (6023A)

2.0Ω (6028A)

load resistor that is capable of safely dissipating 250 watts. Proceed as follows:

a.Connect the test equipment as shown in Figure 2-8.

b.Switch the unit's power on and turn the output voltage all the way up.

c.Turn up output current to: 29.0Vdc (6023A)

10Vdc (6028A)

Check that the unit's CC LED remains lighted.

d.Check that the rms noise current measured by the current probe and rms voltmeter is no more than: 15mA rms (6023A).

5mA rms (6028A)

Figure 2-8. CC PARD Test Setup