Table A-1. Specifications, Option 002 (continued)

Pulse Timing

Low Bias or AC DROPOUT will go false after 5V supply stabilizes.

Bias Supplies

DC Output Ratings: (25°C ± 5)

No Load to Full Load 104V to 127V line.

+5V ± 3% at 100mA

+15V ± 3% at 75mA -15V ± 4% at 75mA

Short Circuit Output Current:

+5V

125mA ± 6%

 

+ 15V

103mA ± 6%

 

-15V

103mA ± 6%

 

PARD (Typical):

 

+ 5V

25mV pk-pk

1.5mV Rms

+ 15V

Same

Same

-15V

Same

Same

Isolation:

Status Indicator lines and Remote Control lines may be floated a maximum of 240Vdc (6010A, 250Vdc, 6011A, 6012B) from ground from the power supply or from each other. These lines may not be connected to any primary circuits.

Jumpers Designation

 

W1--jumpered:

OV indication @ A7J3-17 is active (lo) if OVP; Remote Trip or Remote Inhibit is

 

active.

W1--open:

OV indication is active (lo) if OVP or Remote Trip is active.

Normal operation as shipped:

W3 and W4 jumpered W2 and W5 open.

OVP Programmable

CV: W2 jumpered; W3 open or

 

CC: W5 jumpered; W4 open

 

S1A,B in open position