Chapter 5 Error Messages

Execution Errors

-141

Invalid character data

 

Either the character data element contained an invalid character or the

 

particular element received was not valid for the header.

-144

Character data too long

 

The character data element contained too many characters.

-148

Character data not allowed

 

A discrete parameter was received but a character string or a numeric

 

parameter was expected. Check the list of parameters to verify that you have

 

used a valid parameter type.

 

Example: DISP:TEXT ON

-151

Invalid string data

 

An invalid character string was received. Check to see if you have enclosed

 

the character string in single or double quotes.

 

Example: DISP:TEXT ON

-158

String data not allowed

 

A character string was received but is not allowed for the command. Check

 

the list of parameters to verify that you have used a valid parameter type.

 

Example: TRIG:DEL ‘zero’

-160 to -168

Block data errors

 

The power supply does not accept block data.

-170 to -178

Expression errors

 

The power supply does not accept mathematical expressions.

-211

Trigger ignored

 

A Group Execute Trigger (GET) or *TRG was received but the trigger was

 

ignored. Make sure that the trigger source should be selected to the bus and

 

the trigger subsystem should be initiated by INIT[:IMM] command.

-213

Init ignored

 

An INITiate command was received but could not be executed because a

measurement was already in progress. Send a device clear to halt a measurement in progress and place the power supply in the ‘‘idle’’ state.