4Save the operating state.

Chapter 3 Front-Panel Operation

Storing and Recalling Operating States

Store

4Save the operating state.

The operating state is now stored. To recall the stored state, go to the following steps.

DONE

Recall

5Turn on the recall mode.

Memory location “1” will be displayed in the recall mode.

RECALL 1

This message appears on the display for approximately 3 seconds.

6 Recall the stored operating state.

Turn the knob to the right to change the displayed storage location to 3.

RECALL 3

If this setting is not followed within 3 seconds with Recall key stroke, the power supply returns to normal operating mode and will not recall the instrument state 3 from memory.

Recall

7Restore the operating state.

The power supply should now be configured in the same state as when you stored the state on the previous steps.

DONE

This message appears on the display for approximately 1 second.

Remote interface operation:

*SAV {123} Store an operating state to a specified location

*RCL {123} Recall a previously stored state from a specified location