Chapter 5 Error Messages

Calibration Errors

 

Calibration Errors

 

The following errors indicate failures that may occur during a calibration.

 

Refer to the Service Guide for more information.

701

Cal security disabled by jumper

 

The calibration security feature has been disabled with a jumper inside the

 

power supply. When applicable, this error will occur at power-on to warn you

 

that the power supply is unsecured.

702

Cal secured

 

The power supply is secured against calibration.

703

Invalid secure code

 

An invalid calibration security code was received when attempting to unsecure

 

or secure the power supply. You must use the same security code to unsecure

 

the power supply as was used to secure it, and vice versa. The security code

 

may contain up to 12 alphanumeric characters. The first character must be a

 

letter.

704

Secure code too long

 

A security code was received which contained more than 12 characters.

705

Cal aborted

 

A calibration in progress is aborted when you press any front-panel key, send

 

a device clear, or change the local/remote state of the instrument.

708

Cal output disabled

 

Calibration is aborted by sending OUTP OFF command during calibrating a

 

output.

712

Bad DAC cal data

 

The specified DAC calibration values (CAL:VOLT or CAL:CURR) are out of

 

range. Note that the new calibration constants are not stored in the non-volatile

 

memory.

713

Bad readback cal data

 

The specified readback calibration values (CAL:VOLT or CAL:CURR) are out

of range. Note that the new calibration constants are not stored in the non- volatile memory.