Chapter 3 Front-Panel Operation

Programming Overcurrent Protection

3Clear the overcurrent condition and exit this menu.

Now, when you press OverCurrent key again, the “DONE’’ message is displayed for a second and the OCP annunciator will not blink any more. The output will return to meter mode. The knob is selected for current control.

Notice that the power supply is operated in the constant current (CC) mode.

Adjust OCP trip level

1 Raise the OCP trip level.

Press OverCurrent key and turn the knob to raise the OCP trip level.

2 Move to the OCP CLEAR mode.

OCP CLEAR

Press the key to move to the OCP CLEAR mode. The “OCP ON” message appears on the display. Turn the knob to the right until the above message appears on the display.

3Clear the overcurrent condition and exit this menu.

Now, when you press OverCurrent key again, the “DONE’’ message is displayed for a second and the OCP annunciator will not blink any more. The output will return to the meter mode.

Remote interface operation:

CURR:PROT {<current>MINMAX}

Set the OCP level

CURR:PROT:STAT {OFFON}

Disable or enable the OCP circuit

CURR:PROT:CLE

Clear the tripped OCP circuit