Display
Limit

Chapter 3 Front-Panel Operation

Constant Current Operation

Voltage

1

 

Current

5 Adjust the knob for the desired output current.

 

Check that the Limit annunciator still blinks. Set the knob for current control. The second digit of the ammeter will be blinking. Change the blinking digit using the resolution selection keys and adjust the knob to the desired output current.

Display

 

Limit

6 Return to the meter mode.

 

Press key or let the display time-out after several seconds to return the meter mode. Notice that the Limit annunciator turns off and the display shows “OUTPUT OFF” message.

Output

 

On/Off

7 Enable the output.

 

The OFF annunciator turns off and the 8V* (or 25V**) or 20V* (or 50V**), OVP, OCP and CC annunciators are lit. Notice that the display is in the meter mode. In the meter mode, the display shows the actual output voltage and current.

Refer to “Programming Overvoltage Protection” and “Programming Overcurrent Protection” sections, starting on page 42 and page 46 for more information on OVP and OCP annunciators.

8Verify that the power supply is in the constant current mode.

If you operate the power supply in the constant current (CC) mode, verify that the CC annunciator is lit. If the CV annunciator is lit, choose a higher voltage limit.

N o t e

During actual CC operation, if a load change causes the voltage limit to be exceeded,

 

the power supply will automatically crossover to constant voltage mode at the preset

 

voltage limit and the output current will drop proportionately.

 

 

 

Remote interface operation:

 

 

VOLTage {<voltage>MINMAX}

Set the voltage

 

CURRent {<current>MINMAX}

Set the current

 

OUTPut ON

Enable the output

3

1You can use the resolution selection keys to move the blinking digit to the right or left when setting the current.

*For Agilent E3633A Model **For Agilent E3634A Model

39

Page 40
Image 40
Agilent Technologies E3633A, E3634A manual Adjust the knob for the desired output current