I/O Config

I/O Config

I/O Config

Chapter 3 Front-Panel Operation

Remote Interface Configuration

To Set the GPIB Address

To configure the power supply for the GPIB interface, proceed as follows:

1 Turn on the remote configuration mode.

GPIB / 488

You will see the above message on the front-panel display if the power supply has not been changed from the factory setting. If “RS-232” appears, choose “GPIB / 488” by turning the knob to the right.

2 Move to the GPIB address setting mode.

ADDR 05

The address is set to “05” when the power supply is shipped from the factory. Notice that a different GPIB address may appear if the power supply has been changed from the factory setting.

3Turn the knob to change the GPIB address.

The displayed address is changed when turning the knob to the right or left.

4Save the change and turn off the I/O configuration mode.

CHANGE SAVED

 

The address is stored in non-volatile memory, and does not change when power

 

has been off or after a remote interface reset. The power supply displays a

 

message to show that the change is now in effect. If the GPIB address is not

 

changed, “NO CHANGE” will be displayed for one second.

 

 

N o t e

To exit the I/O configuration mode without any further changes, press the “I/O

 

Config” key until the “NO CHANGE” message is displayed.

 

 

60

Page 61
Image 61
Agilent Technologies E3634A, E3633A manual To Set the Gpib Address, Turn on the remote configuration mode