Agilent Technologies E3633A, E3634A The following program segments show how to use the VOLT UP or

Models: E3634A E3633A

1 175
Download 175 pages 17.89 Kb
Page 88
Image 88
The following program segments show how to use the VOLT UP or

VOLTage Example

Chapter 4 Remote Interface Reference

Output Setting and Operation Commands

VOLTage {<voltage> MINimum MAXimumUPDOWN}

This command programs the immediate voltage level of the power supply. The immediate level is the voltage value of the output terminals.

The VOLTage command changes the output of the power supply to the newly programmed value regardless of the output range presently selected.

You can substitute ‘‘MINimum’’ or ‘‘MAXimum’’ in place of a specific value for the voltage parameter. MIN selects the lowest voltage values of ‘‘0’’ volts. MAX selects the highest voltage values allowed for the selected range.

This command also increases or decreases the immediate voltage level using the ‘‘UP’’ or ‘‘DOWN’’ parameter by a predetermined amount. The command VOLTage:STEP sets the amount of increase or decrease. Notice that a new increment setting will cause an execution error -222 (Data out of range) when the maximum or the minimum rated voltage is exceeded.

The following program segments show how to use the VOLT UP or

VOLT DOWN command to increase or decrease the output voltage with the VOLT:STEP command.

‘‘VOLT:STEP 0.01’’

Set the step size to 0.01 V

‘‘VOLT UP’’

Increase the output voltage

‘‘VOLT:STEP 0.02’’

Set the step size to 0.02 V

‘‘VOLT DOWN’’

Decrease the output voltage

VOLTage? [MINimum MAXimum]

This query returns the presently programmed voltage level of the power supply. VOLT? MAX and VOLT? MIN return the highest and lowest programmable voltage levels for the selected range.

87