Agilent Technologies E3633A, E3634A manual Triggering Commands

Models: E3634A E3633A

1 175
Download 175 pages 17.89 Kb
Page 94
Image 94
Triggering Commands

Chapter 4 Remote Interface Reference

Triggering Commands

Triggering Commands

INITiate

This command causes the trigger system to initiate. This command completes one full trigger cycle when the trigger source is an immediate and initiates the trigger subsystem when the trigger source is bus.

TRIGger:DELay {<seconds> MINimum MAXimum}

This command sets the time delay between the detection of an event on the specified trigger source and the start of any corresponding trigger action on the power supply output. Select from 0 to 3600 seconds. MIN = 0 seconds.

MAX = 3600 seconds. At *RST, this value is set to 0 seconds.

TRIGger:DELay?

This command queries the trigger delay.

TRIGger:SOURce {BUS IMMediate}

This command selects the source from which the power supply will accept a trigger. The power supply will accept a bus (software) trigger or an internal immediate trigger. At *RST, the bus trigger source is selected.

TRIGger:SOURce?

This command queries the present trigger source. Returns ‘‘BUS’’ or ‘‘IMM’’.

*TRG

This command generates a trigger to the trigger subsystem that has selected a bus (software) trigger as its source (TRIG:SOUR BUS). The command has the same effect as the Group Execute Trigger (GET) command. For RS-232 operation, make sure the power supply is in the remote interface mode by sending the SYST:REM command first.

93

4

Page 94
Image 94
Agilent Technologies E3633A, E3634A manual Triggering Commands