Image 91
This query returns the currently selected range. The returned parameter for the Agilent E3633A is ‘‘P20V’’ (HIGH) or ‘‘P8V’’ (LOW) and the parameter for the Agilent E3634A is ‘‘P50V’’ (HIGH) or ‘‘P25V’’ (LOW).

Chapter 4 Remote Interface Reference

Output Setting and Operation Commands

VOLTage:RANGe?

This query returns the currently selected range. The returned parameter for the Agilent E3633A is ‘‘P20V’’ (HIGH) or ‘‘P8V’’ (LOW) and the parameter for the Agilent E3634A is ‘‘P50V’’ (HIGH) or ‘‘P25V’’ (LOW).

MEASure:CURRent?

This command queries the current measured across the current sense resistor inside the power supply.

MEASure[:VOLTage]?

This command queries the voltage measured at the sense terminals of the power supply.