Chapter 4 Remote Interface Reference

Using the APPLy Command

Using the APPLy Command

The APPLy command provides the most straightforward method to program the power supply over the remote interface. You can select the output voltage and current in one command.

APPLy {<voltage> DEF MIN MAX}[,{<current> DEF MIN MAX}]

This command is combination of VOLTage and CURRent commands. As long as the newly programmed values are within the presently selected range, the output voltage and current are changed as soon as the command is executed.

The APPLy command changes the power supply’s output to the newly programmed values only if the programmed values are valid within the presently selected range. An execution error will occur if the programmed values are not valid within the selected range.

You can substitute ‘‘MINimum’’, ‘‘MAXimum’’, or ‘‘DEFault’’ in place of a specific value for the voltage and current parameters. MIN selects the lowest values of ‘‘0’’ volts and ‘‘0’’ amps. MAX selects the highest values allowed for the selected range.

The default values of voltage and current are ‘‘0’’ volts and ‘‘20’’* or ‘‘7’’** amps regardless of the presently selected range. For more details of parameters, see Table 4-1 for the Agilent E3633A model and Table 4-2 for the Agilent E3634A model.

If you specify only one parameter of the APPLy command, the power supply regards it as voltage setting value.

APPLy?

This command queries the power supply’s present voltage and current setting values and returns a quoted string. The voltage and current are returned in sequence as shown in the sample string below (the quotation marks are returned as part of the string).

‘‘8.00000,20.00000’’* or ‘‘25.00000,7.00000’’**

In the above string, the first number 8.00000 is the voltage setting value and the second number 20.00000 is the current setting value.

4

*For Agilent E3633A Model **For Agilent E3634A Model