executed due to the current device state.

Chapter 5 Error Messages

Execution Errors

-221

Settings conflict

 

Indicates that a legal program data element was parsed but could not be

 

executed due to the current device state.

-222

Data out of range

 

A numeric parameter value is outside the valid range for the command.

 

Example: TRIG:DEL -3

-223

Too much data

 

A character string was received but could not be executed because the string

 

length was more than 40 characters. This error can be generated by the

 

CALibration:STRing command.

-224

Illegal parameter value

 

A discrete parameter was received which was not a valid choice for the

 

command. You may have used an invalid parameter choice.

 

Example: DISP:STAT XYZ (XYZ is not a valid choice).

-330

Self-test failed

 

The power supply’s complete self-test failed from the remote interface (*TST?

 

command). In addition to this error, more specific self-test errors are also

 

reported. See also ‘‘Self-Test Errors’’, starting on page 130.

-350

Too many errors

 

The error queue is full because more than 20 errors have occurred. No

 

additional errors are stored until you remove errors from the queue. The error

 

queue is cleared when power has been off, or after a *CLS (clear status)

 

command has been executed.

-410

Query INTERRUPTED

 

A command was received which sends data to the output buffer, but the output

 

buffer contained data from a previous command (the previous data is not

 

overwritten). The output buffer is cleared when power has been off, or after a

 

*RST (reset) command has been executed.

-420

Query UNTERMINATED

 

The power supply was addressed to talk (i.e., to send data over the interface)

 

but a command has not been received which sends data to the output buffer.

For example, you may have executed an APPLy command (which does not generate data) and then attempted an ENTER statement to read data from the remote interface.