Agilent Technologies E3634A, E3633A manual OVP and OCP Accuracy, ±% of output + offset

Models: E3634A E3633A

1 175
Download 175 pages 17.89 Kb
Page 165
Image 165
OVP and OCP Accuracy, ±(% of output + offset)

Chapter 8 Specifications

Performance Specifications

Transient Response Time

Less than 50 msec for output to recover to within 15 mV following a change in output current from full load to half load or vice versa

Command Processing Time

Average time for output voltage to begin to change after receipt of digital data when the power supply is connected directly to the GPIB or RS-232 is less than 100 msec

OVP and OCP Accuracy, ±(% of output + offset)

OVP

0.5% + 0.5 V

OCP

0.5% + 0.5 A

Activation time : Average time for output to start to drop after OVP or OCP condition occurs.

OVP

<1.5 msec when the trip voltage is equal or greater than 3 V

 

<10 msec when the trip voltage is less than 3 V

OCP

<10 msec

164

Page 165
Image 165
Agilent Technologies E3634A, E3633A manual OVP and OCP Accuracy, ±% of output + offset