Chapter 5 Error Messages

Calibration Errors

714

 

Bad OVP cal data

 

 

The overvoltage protection calibration constant is out of range. Note that the

 

 

new calibration constants are not stored in the non-volatile memory.

715

 

Bad OCP cal data

 

 

The overcurrent protection calibration constant is out of range. Note that the

 

 

new calibration constants are not stored in the non-volatile memory.

716

 

Bad DAC DNL error correction data

 

 

Invalid data measured during the calibration for DAC differential nonlinearity

 

 

error correction.

717

 

Cal OVP or OCP status enabled

 

 

Overvoltage protection status or overcurrent protection status is enabled. You

 

 

must set both overvoltage and overcurrent protection status to OFF before and

 

 

during the calibration.

718

 

Gain out of range for Gain Error Correction

 

 

The slope of the DAC gain is out of range. Hardware fails.

740

 

Cal checksum failed, secure state

741

 

Cal checksum failed, string data

742

 

Cal checksum failed, store/recall data in location 0

743

 

Cal checksum failed, store/recall data in location 1

744

 

Cal checksum failed, store/recall data in location 2

745

 

Cal checksum failed, store/recall data in location 3

746

 

Cal checksum failed, DAC cal constants

747

 

Cal checksum failed, readback cal constants

748

 

Cal checksum failed, GPIB address

749

 

Cal checksum failed, internal data

750

 

Cal checksum failed, DAC DNL error correction data

 

 

 

 

132

Page 133
Image 133
Agilent Technologies E3634A, E3633A manual Error Messages Calibration Errors 714 Bad OVP cal data, Bad OCP cal data