Chapter 8 Specifications

Performance Specifications

Performance Specifications

Table 1-1. Performance Specifications

Parameter

 

Agilent E3633A

Agilent E3634A

 

 

 

 

Output Ratings

Low Range

0 to +8 V/0 to 20 A

0 to +25 V/0 to 7 A

(@ 0 °C - 40 °C)

 

 

 

High Range

0 to +20 V/0 to 10 A

0 to +50V/0 to 4 A

 

 

 

 

 

Programming Accuracy[1]

Voltage

0.05% + 10 mV

12 months (@ 25 °C ± 5 °C),

 

 

 

Current

0.2% + 10 mA

±(% of output + offset)

 

 

 

 

 

 

Readback Accuracy[1]

Voltage

0.05% + 5 mV

12 months (over GPIB and

 

 

 

Current

0.15% + 5 mA

RS-232 or front panel with

 

 

 

respect to actual output @ 25 °C

 

 

 

± 5 °C), ±(% of output + offset)

 

 

 

 

 

 

 

Ripple and Noise

Normal mode

<0.35 mV rms and

<0.5 mV rms and

(with outputs ungrounded, or

voltage

3 mV p-p

3 mV p-p

with either output terminal

 

 

 

Normal mode

<2 mA rms

grounded, 20 Hz to 20 MHz)

current

 

 

 

 

 

 

 

 

 

Common mode

<1.5 uA rms

 

current

 

 

 

 

 

Load Regulation,

Voltage

<0.01% + 2 mV

±(% of output + offset)

 

 

 

Current

<0.01% + 250 uA

 

 

 

 

Line Regulation,

Voltage

<0.01% + 2 mV

±(% of output + offset)

 

 

 

Current

<0.01% + 250 uA

 

 

 

 

 

Programming Resolution

Voltage

1 mV

3 mV

 

 

 

 

 

Current

1 mA

0.5 mA

 

 

 

 

Readback Resolution

Voltage

0.5 mV

1.5 mV

 

 

 

 

 

Current

1 mA

0.5 mA

 

 

 

 

Front Panel Resolution

Voltage

1 mV

 

 

 

 

Current

1 mA (< 10A), 10mA (³ 10A)

 

 

 

 

[1]Accuracy specifications are after an 1-hour warm-up with no load and calibration at 25 °C.

163

8

Page 164
Image 164
Agilent Technologies E3633A, E3634A manual Specifications Performance Specifications, 163