Agilent Technologies N6700 manual Connecting the Outputs, Screws securely tightened

Models: N6700

1 109
Download 109 pages 3.15 Kb
Page 25
Image 25

Installation 2

Connecting the Outputs

 

SHOCK HAZARD Turn off AC power before making rear panel connections.

WARNING

All wires and straps must be properly connected with the terminal block

 

 

screws securely tightened.

 

 

 

Disconnect the connector plug to make your wire connections. The

 

12A connector plug accepts wires sizes from AWG 12 to AWG 30. The

 

20A connector plug accepts wires sizes from AWG 10 to AWG 24. The

 

50A connector plug accepts wires sizes from AWG 6 to AWG 20. Wire

 

sizes smaller than AWG 20 are not recommended. Connect the load

 

wires to the + and - terminals. Connect the sense wires to the +s and

 

-s terminals. Sense jumpers are provided for local sensing.

 

Securely fasten all wires by tightening the screw terminals. Insert the

 

connector plug into the back of the unit. Secure the 12 A connector

 

by tightening the locking screws. A chassis ground binding post is

 

located next to the AC input connector for ground connections.

 

On power modules with the 50A sense connector, the +LS and –LS terminals

CAUTION

 

are ONLY used for local sense connections as illustrated. Do not connect these

 

terminals in any other way.

 

 

 

 

 

 

 

TIGHTEN

 

 

 

 

 

SCREWS

 

 

 

 

20 A CONNECTOR

 

 

 

12 A

 

 

 

 

LOCKING

CONNECTOR

+S

+

-S

LOCKING

 

 

SCREW

+S

+

-S

 

 

SCREW

 

 

 

 

INSERT WIRES

 

SENSE JUMPERS

 

INSERT WIRES

 

 

 

 

TWIST LEADS

 

 

 

 

 

INSTALLED FOR

 

 

 

 

 

 

LOCAL SENSING

 

TWIST LEADS

 

 

 

 

 

 

 

+

 

+

 

 

 

 

LOAD

 

LOAD

 

 

50 A

CONNECTOR

+

+

LOAD

50A

SENSE

+S +LS -LS -S

SENSE JUMPERS INSTALLED FOR LOCAL SENSING

Series N6700 User’s Guide

25

Page 25
Image 25
Agilent Technologies N6700 manual Connecting the Outputs, Screws securely tightened, Terminals in any other way