Appendix A Specifications

Agilent Models N6731B - N6736B and N6741B - N6746B

Performance Specifications

 

N6731B/

N6732B/

N6733B/

N6734B/

N6735B/

N6736B/

 

N6741B

N6742B

N6743B

N6744B

N6745B

N6746B

 

 

 

 

 

 

 

DC Output Ratings:

 

 

 

 

 

 

Voltage

5 V

8 V

20 V

35 V

60 V

100 V

 

 

NOTE 2

 

 

 

 

Current NOTE 1

10 A / 20 A

6.25 A / 12.5 A

2.5 A / 5 A

1.5 A / 3 A

0.8 A / 1.6 A

0.5 A / 1 A

Power

50 W / 100 W

50 W / 100 W

50 W / 100 W

52.5W / 105W

50 W / 100 W

50 W / 100 W

 

 

 

 

 

 

Output Ripple and Noise (PARD):

 

 

 

 

 

(from 20 Hz – 20 MHz)

 

 

 

 

 

 

CV peak-to- peak

10 mV / 11mV

12 mV

14 mV

15 mV

25 mV

30 mV

CV rms

2 mV

2 mV

3 mV

5 mV

9 mV

18 mV

 

 

 

 

 

 

Load Effect (Regulation):

 

 

 

 

 

(with output change from no load to full load, up to a maximum load-lead drop of 1V/lead)

 

Voltage

5 mV

6 mV

9 mV

11 mV

13 mV / 16 mV

20 mV / 30 mV

Current

2 mA

2 mA

2 mA

2 mA

2 mA

2 mA

 

 

 

 

 

 

Source Effect (Regulation):

 

 

 

 

 

Voltage

1 mV

2 mV

2 mV

4 mV

6 mV

10 mV

Current

1 mA

1 mA

1 mA

1 mA

1 mA

1 mA

Programming Accuracy:

(@ 23 °C ±5 °C after 30 minute warm-up. Applies from minimum to maximum programming range)

Voltage

0.1% + 19 mV

0.1% + 19 mV

0.1% + 20 mV

0.1% + 35 mV

0.1% + 60 mV

0.1% +100 mV

Current

0.15% + 20 mA

0.15% + 20 mA

0.15% + 20 mA

0.15% + 20 mA

0.15% + 20 mA

0.15% + 10mA

Measurement Accuracy: (at 23 °C ±5 °C)

Voltage

0.1% + 20 mV

0.1% + 20 mV

0.1% + 20 mV

0.1% + 35 mV

0.1% + 60 mV

0.1% +100 mV

Current

0.15% + 20 mA

0.15% + 10 mA

0.15% + 5 mA

0.15% + 4 mA

0.15% + 4 mA

0.15% + 2 mA

Load Transient Recovery Time:

(time to recover to within the settling band following a load change from 50% to 100% and from 100% to 50% of full load.)

 

NOTE 3

NOTE 3

 

 

 

 

Voltage settling band

±0.08 V / 0.1 V

±0.08 V / 0.1 V

± 0.2 V / 0.3 V

± 0.2 V / 0.3 V

± 0.4 V / 0.5 V

± 0.5 V / 1.0 V

Time

< 200 s

< 200 s

< 200 s

< 200 s

< 200 s

< 200 s

 

 

 

 

 

 

 

1Output current is derated 1% per °C above 40°C.

2When relay option 760 is installed on Model N6742B, the maximum output current will be limited to 10 A.

3When relay option 760 or 761 is installed, the settling band is ±0.10V/0.125 V. Option 760 is not available on Model N6741B.