Agilent Technologies manual Agilent N6700B, N6701A, N6702A MPS Mainframes

Models: N6700

1 109
Download 109 pages 3.15 Kb
Page 86
Image 86

Appendix A Specifications

Agilent N6700B, N6701A, N6702A MPS Mainframes

Supplemental Characteristics

 

N6700B, N6701A, N6702A

Command Processing Time:

 

 

≤ 1 ms from receipt of command to start of output change

Protection Response Characteristics:

INH input

5 s from receipt of inhibit to start of shutdown

Fault on coupled outputs

< 10 s from receipt of fault to start of shutdown

Digital Control Characteristics:

Maximum voltage ratings

Pins 1 and 2 as FLT output

Pins 1 - 7 as digital/trigger outputs (pin 8 = common)

Pins 1 - 7 as digital/trigger inputs and pin 3 as INH input (pin 8 = common)

+16.5 VDC/5 VDC between pins

(pin 8 is internally connected to chassis ground).

Maximum low-level output voltage = 0.5 V @ 4 mA Maximum low-level sink current = 4 mA

Typical high-level leakage current = 1 mA @ 16.5 VDC

Maximum low-level output voltage = 0.5 V @ 4 mA; 1 V @ 50 mA; 1.75 V @ 100 mA

Maximum low-level sink current = 100 mA

Typical high-level leakage current = 0.8 mA @ 16.5 VDC

Maximum low-level input voltage = 0.8 V Minimum high-level input voltage = 2 V

Typical low-level current = 2 mA @ 0 V (internal 2.2k pull-up) Typical high-level leakage current = 0.12 mA @ 16.5 VDC

Interface Capabilities:

GPIB

SCPI - 1993, IEEE 488.2 compliant interface

LXI Compliance

Class C (only applies to units with LXI label on front panel)

USB 2.0

Requires Agilent IO Library version M.01.01 or 14.0 and up

10/100 LAN

Requires Agilent IO Library version L.01.01 or 14.0 and up

Built-in Web server

Requires Internet Explorer 5+ or Netscape 6.2+

Regulatory Compliance:

EMC

Complies with EMC directive for Class A test and

 

measurement products.

 

Complies with Australian standard and carries C-Tick mark.

 

This ISM device complies with Canadian ICES-001.

 

Cet appareil ISM est conforme à la norme NMB-001 du

 

Canada.

 

Electrostatic discharges greater than 1 kV near the I/O

 

connectors may cause the unit to reset and require

 

operator intervention.

Safety

Complies with European Low Voltage Directive and carries

 

the CE-marking.

 

Complies with US and Canadian safety standards for test

 

and measurement products.

86

Series N6700 User’s Guide