Installation 2

Remote Sense Connections

WARNING SHOCK HAZARD Turn off AC power before making or changing rear panel connections.

Remote sensing improves the voltage regulation at the load by monitoring the voltage there instead of at the output terminals. This allows the power system to automatically compensate for the voltage drop in the load leads. Remote sensing is especially useful for CV operation with load impedances that vary or have significant lead resistance. It has no effect during CC operation. Because sensing is independent of other power system functions, remote sensing can be used regardless of how the power system is programmed.

12A & 20 A

+S

+

-S

CONNECTOR

TWIST

 

 

 

PAIR

 

 

TWIST

 

 

 

 

 

 

LEADS

 

+

LOAD

 

 

 

 

+

+

LOAD

50A CONNECTOR

50A SENSE

+S +LS -LS -S

TWIST

PAIR

 

Connect the unit for remote sensing by first disconnecting the straps

 

between sense and load terminals. Make your connections as shown

 

in the following figure. Connect the sense leads as close to the load as

 

possible. Refer to the “Wire Size” section for information about

 

selecting the proper wire size. Best results are obtained by using the

 

shortest load leads practical. It is recommended to keep load leads

 

under 14.7 meters (50 feet) per lead because of inductance effects.

 

The sense leads carry only a few milliamperes of current and can be

 

a lighter gauge than the load leads. However, note that any voltage

 

drop in the sense leads can degrade the voltage regulation of the

 

instrument. Try to keep the sense lead resistance less than about

 

0.5Ω per lead (this requires 20 AWG or heavier for a 50 foot length).

 

When remote sensing on power modules with the 50A sense connector, do not

CAUTION

 

connect the +LS and –LS terminals. These are dedicated for local sensing only.

 

 

Series N6700 User’s Guide

29

Page 29
Image 29
Agilent Technologies N6700 manual Remote Sense Connections, Drop in the sense leads can degrade the voltage regulation