Agilent Technologies N6700 manual 20 mA

Models: N6700

1 109
Download 109 pages 3.15 Kb
Page 85
Image 85

Specifications Appendix A

Supplemental Characteristics

 

N6773A

N6774A

N6775A

N6776A

 

 

 

 

Programming Ranges:

 

 

 

Voltage

30 mV – 20.4 V

40 mV – 35.7 V

70 mV – 61.2 V

100 mV – 102 V

Current

30 mA – 15.3 A

15 mA – 8.67 A

7.5 mA – 5.1 A

4.5 mA – 3.06 A

 

 

 

 

Programming Resolution:

 

 

 

Voltage

7 mV

10 mV

18 mV

28 mV

Current

9 mA

6 mA

3 mA

1.5 mA

Measurement Resolution:

 

 

 

Voltage

10 mV

18 mV

30 mV

50 mV

Current

9 mA

6 mA

3 mA

1.5 mA

Temperature Coefficient per °C:

 

 

 

Voltage Programming

0.01% + 0.2 mV

0.01% + 0.5 mV

0.01% + 0.5 mV

0.01% + 1 mV

Current Programming

0.01% + 0.5 mA

0.01% + 0.5 mA

0.01% + 0.1 mA

0.01% + 0.1 mA

Voltage Measurement

0.01% + 0.2 mV

0.01% + 0.2 mV

0.01% + 0.5 mV

0.01% + 0.5 mV

Current Measurement

0.01% + 0.5 mA

0.01% + 0.5 mA

0.01% + 0.05 mA

0.01% + 0.05 mA

Output Ripple and Noise (PARD):

 

 

 

CC rms

6 mA

6 mA

6 mA

6 mA

 

 

 

 

Common Mode Noise:

 

 

 

(from 20 Hz – 20 MHz; from either output to chassis)

 

 

 

Rms

2 mA

2 mA

2 mA

2 mA

Peak-to- peak

< 20 mA

< 20 mA

< 20 mA

< 20 mA

Over-voltage Protection:

 

 

 

Accuracy

0.25% +100 mV

0.25% + 130 mV

0.25% + 260 mV

0.25% + 650 mV

Accuracy w/opt 760

0.25% + 700 mV

0.25% + 700 mV

0.25% + 400 mV

0.25% + 650 mV

Accuracy w/opt 761

0.25% + 500 mV

0.25% + 350 mV

0.25% + 350 mV

0.25% + 650 mV

Maximum setting

22 V

38.5 V

66 V

110 V

Response time

50 s from occurrence of over-voltage condition to start of output shutdown

 

 

Remote Sense Capability:

Outputs can maintain specifications with up to a 1-volt drop per load lead.

Series and Parallel Operation:

Identically rated outputs can be operated directly in parallel or can be connected for straight series operation. Auto- series and auto-parallel operation is not available.

Maximum Up-programming and Down-programming Time with full resistive load: (time from 10% to 90% of total voltage excursion)

From 0 V to full scale

20 ms

20 ms

20 ms

20 ms

and full scale to 0 V

 

 

 

 

Maximum Up-programming and Down-programming Settling Time with full resistive load:

(time from start of voltage change until voltage settles within 0.1% of the full-scale voltage of its final value)

From 0 V to full scale

100 ms

100 ms

100 ms

100 ms

and full scale to 0 V

 

 

 

 

Series N6700 User’s Guide

85

Page 85
Image 85
Agilent Technologies N6700 manual 20 mA