Installation 2

OUTPUT 2

+S +

 

 

 

-S

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SENSE

JUMPERS

INSTALLED

OUTPUT 1

+S

+

-S

SENSE

JUMPERS

INSTALLED

OUTPUT 2

+S +

 

-S

 

SENSE JUMPER

INSTALLED

OUTPUT 1

+S

+

-S

TWIST LEADS

+

LOAD

TWIST LEADS

+

LOAD

WITH LOCAL SENSING

WITH REMOTE SENSING

The following figure shows the connections for 50A power modules.

OUTPUT 3

+

+S +LS -LS -S

OUTPUT 1

+

+S +LS -LS -S

OUTPUT 3

+

+S +LS -LS -S

OUTPUT 1

+

+S +LS -LS -S

SENSE

SENSE

JUMPER

INSTALLED

JUMPERS

INSTALLED

TWIST LEADS

+

+

TWIST LEADS

LOAD

WITH LOCAL SENSING

Setting the Outputs

LOAD

WITH REMOTE SENSING

 

Outputs connected together in series cannot be grouped.

 

To program outputs connected in series, first program the current

 

limit of each output to the total desired current limit point. Then

 

program the voltage of each output so that the sum of both voltages

 

equals the total desired operating voltage. The simplest way to

 

accomplish this is to program each output to one half of the total

 

desired operating voltage.

 

The operating mode of each output is determined by the output’s programmed

NOTE

settings, operating point, and load condition. Because these conditions may

 

 

change during series operation, the front panel status indicator will reflect

 

these changes. This is normal. Momentary status changes are also normal.

 

 

Series N6700 User’s Guide

33

Page 33
Image 33
Agilent Technologies N6700 manual Setting the Outputs, Desired operating voltage