3 Getting Started

Connecting to the Interfaces

 

Electrostatic discharges greater than 1 kV near the interface connectors may

CAUTION

cause the unit to reset and require operator intervention.

 

 

 

 

The Agilent N6700 MPS supports GPIB, LAN, and USB interfaces. All

 

three interfaces are live at power-on. The front panel IO indicator

 

comes on whenever there is activity on the remote interfaces.

GPIB Interface

 

 

 

 

For detailed information about GPIB interface connections, refer to the Agilent

NOTE

 

 

Technologies USB/LAN/GPIB Interfaces Connectivity Guide, located on the

 

 

 

 

 

 

 

 

Automation-Ready CD that is shipped with your product.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The following steps will help you quickly get started connecting your

 

 

 

 

instrument to the General Purpose Interface Bus (GPIB). The

 

 

 

 

following figure illustrates a typical GPIB interface system.

 

 

 

 

 

 

 

 

 

 

 

GPIB Cable

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PC

Instrument

Instrument

 

 

Connect to GPIB Interface

Connect to GPIB

 

Card installed in PC.

port on instrument.

 

1If you have not already done so, install the Agilent IO Libraries Suite from the Automation-Ready CD that is shipped with your product.

2If you do not have a GPIB interface card installed on your computer, turn off your computer and install the GPIB card.

3Connect your instrument to the GPIB interface card using a GPIB interface cable.

4Use the Connection Expert utility of the Agilent IO Libraries Suite to configure the installed GPIB interface card’s parameters.

5The power system is shipped with its GPIB address set to 5. Use the front panel menu if you need to change the GPIB address.

a Press the Menu key, then use the navigation keys to select System\IO\GPIB.

b Use the numeric keys to enter a new value. Valid addresses are from 0 to 30. Press the Enter key to enter the value. Press the Meter key to exit the menu.

6You can now use Interactive IO within the Connection Expert to communicate with your instrument, or you can program your instrument using the various programming environments.

40

Series N6700 User’s Guide

Page 40
Image 40