Agilent Technologies N6700 Select the Measurement Trigger Source, Sensfunc VOLT,@14, Transientn

Models: N6700

1 109
Download 109 pages 3.15 Kb
Page 73
Image 73

 

 

 

Operating the Power System 4

 

 

To trigger measurements on models that do not have simultaneous

 

 

voltage and current measurement capability, select the measurement

 

 

function as follows.

 

 

 

 

 

 

 

Front Panel:

SCPI Command:

 

 

Not Available

To select the measurement function:

 

 

 

SENS:FUNC ”VOLT”,(@1:4)

 

 

 

SENS:FUNC ”CURR”,(@1:4)

 

 

 

 

 

If a model has simultaneous voltage and current measurements

 

 

capability, then BOTH voltage and current are acquired on any

 

 

triggered measurement, regardless of the setting of the

 

 

SENSe:FUNCtion command.

 

 

 

Select the Measurement Trigger Source

 

 

An immediate trigger command over the bus will generate an immediate trigger

NOTE

 

 

regardless of the trigger source.

 

 

 

 

 

 

Unless you are using a TRIG:ACQ command to trigger the

 

 

measurement, select a trigger source from the following:

Bus

Selects GPIB device trigger, *TRG, or <GET> (Group Execute Trigger).

Pin<n>

Selects a pin on the external port connector as the trigger source.

 

 

<n> specifies the pin number. The selected pin must be configured as a

 

 

Trigger Input in order to be used as a trigger source (see Appendix B).

Transient<n>

Selects the output channel’s transient system as the trigger source.

 

 

<n> specifies the channel. When you select a channel, you must also set up

 

 

that channel’s transient system to generate a trigger out signal. Refer to

 

 

“Generate Trigger Out Signals” and “Program an Arbitrary List” earlier in this

 

 

chapter.

 

 

 

Use the following commands to select a trigger source:

 

 

 

 

 

 

Front Panel:

SCPI Command:

 

 

Not Available

To select Bus triggers for output 1:

 

 

 

TRIG:ACQ:SOUR BUS,(@1)

 

 

 

To select Digital pin triggers:

TRIG:ACQ:SOUR PIN<n>,(@1) where n is the pin number.

To select Transient output triggers:

TRIG:ACQ:SOUR TRAN<n>,(@1) where n is the output channel that will generate the trigger signal.

Series N6700 User’s Guide

73

Page 73
Image 73