4 Operating the Power System

You can also specify a delay to prevent momentary CV-to-CC status changes from tripping the over-current protection.

Front Panel:

SCPI Command:

Select Protect\OCP.

To specify a 10 millisecond delay:

Enter a delay value and press Select.

OUTP:PROT:DEL 0.01,(@1,2)

Check "Start delay on CC" to start the

To start the delay timer by ANY

delay timer by ANY output transition

output transition into CC mode:

into CC mode. Otherwise, the delay

CURR:PROT:DEL:STAR CCTR,

timer will only be started by a settings

(@1,2)

To start the delay timer by a settings

change in voltage, current, or output

state.

change in voltage current or output:

 

CURR:PROT:DEL:STAR SCH,

 

(@1,2)

 

 

Couple Output Protection

Protection coupling lets you disable all output channels when a protection condition occurs on a single output channel. To couple output protection:

Front Panel:

SCPI Command:

Select Protect\Couple.

To enable output protection coupling:

Check the Enable Coupling box and

OUTP:PROT:COUP ON

press Select.

 

Clear Output Protection Functions

If an over-voltage, over-current, over-temperature, power-fail condition, power-limit condition, protection condition, or inhibit signal occurs, the power system turns off the affected output channel. The appropriate operating status indicator on the front panel will be on (e.g. OV, OC, OT, PF, CP+, PROT, INH). To clear the protection function and restore normal operation, first remove that condition that caused the protection fault. Then, clear the protection function as follows:

Front Panel:

SCPI Command:

Select Protect\Clear.

To clear a protection fault on output 1:

Select the Clear button.

OUTP:PROT:CLE(@1)

60

Series N6700 User’s Guide

Page 60
Image 60