Appendix A Specifications

Supplemental Characteristics (continued)

 

N6751A / N6752A

N6753A

N6754A

N6761A / N6762A

 

 

 

 

Up-programming Time with full resistive load:

 

 

 

(time from 10% to 90% of total voltage excursion)

 

 

 

Small voltage step

0 V to 10 V

0 V to 6 V

0 V to 15 V

0 V to 10 V

Time

0.2 ms

0.4 ms

0.35 ms

0.6 ms

Large voltage step

0 V to 50 V

0 V to 20 V

0 V to 60 V

0 V to 50 V

Time

1.5 ms

1.5 ms

2 ms

2.2 ms

Up-programming Settling Time with full resistive load:

 

 

 

(time from start of voltage change to output voltage within 0.1% of full scale value)

 

 

Small voltage step

0 V to 10 V

0 V to 6 V

0 V to 15 V

0 V to 10 V

Time

0.5 ms

0.8 ms

0.8 ms

0.9 ms

Large voltage step

0 V to 50 V

0 V to 20 V

0 V to 60 V

0 V to 50 V

Time

4 ms

3 ms

4.2 ms

4 ms

Down-programming Time with no load:

 

 

 

 

(time from start of voltage change to output voltage < 0.5 V)

 

 

 

Small voltage step

10 V to 0 V

6 V to 0 V

15 V to 0 V

10 V to 0 V

Time

0.3 ms

0.55 ms

0.6 ms

0.3 ms

Large voltage step

50 V to 0 V

20 V to 0 V

60 V to 0 V

50 V to 0 V

Time

1.3 ms

1.8 ms

2.2 ms

1.3 ms

Down-programming Settling Time with no load:

(time from start of voltage change to output voltage within 0.1% of full scale value)

 

 

Small voltage step

10 V to 0 V

6 V to 0 V

15 V to 0 V

10 V to 0 V

Time

0.45 ms

0.8 ms

0.8 ms

0.45 ms

Large voltage step

50 V to 0 V

20 V to 0 V

60 V to 0 V

50 V to 0 V

Time

1.4 ms

2 ms

2.3 ms

1.4 ms

Down-programming Time with Capacitive load:

(time from start of voltage change to output voltage < 0.5 V)

 

 

 

Small voltage step

10 V to 0 V

6 V to 0 V

15 V to 0 V

10 V to 0 V

Time

2.1 ms

2.2 ms

2.3 ms

4.5 ms

Large voltage step

50 V to 0 V

20 V to 0 V

60 V to 0 V

50 V to 0 V

Time

11 ms

8.5 ms

10 ms

23 ms

Capacitive load

NOTE 2

NOTE 3

NOTE 4

NOTE 2

 

1000 F

4700 F

680 F

1000 F

Down-programming Capability:

 

 

 

 

Continuous power

7 W

12.5 W

12.5 W

7 W

Peak current

7 A

15 A

6 A

3.8 A

High-Speed Test Extensions:

High-speed test extensions are included in Models N676xA and are available as Option 054 in Models N675xA.

Maximum number of List steps = 512 Maximum List dwell time in seconds = 262 Maximum List repetitions = 256, or infinite Maximum measurement points = 4096 Maximum sample rate = 50 kHz

2Modules can discharge a 1000∝F capacitor from full scale to 0V at a rate of 4 times/second.

3Modules can discharge a 4700∝F capacitor from full scale to 0V at a rate of 4 times/second.

4Modules can discharge a 680∝F capacitor from full scale to 0V at a rate of 4 times/second.

78

Series N6700 User’s Guide

Page 78
Image 78
Agilent Technologies N6700 manual Up-programming Time with full resistive load, Down-programming Settling Time with no load