Appendix A Specifications

Agilent Models N6751A/N6752A, N6753A/N6754A, N6761A/N6762A

Performance Specifications

 

N6751A / N6752A

N6753A

N6754A

N6761A / N6762A

 

 

 

 

 

DC Output Ratings:

 

 

 

 

Voltage

50 V

20 V

60 V

50 V

Current (derated 1% per °C above 40°C)

5 A / 10A

50 A

20 A

1.5 A / 3 A

Power

50 W / 100 W

300 W

300 W

50 W / 100 W

Output Ripple and Noise (PARD):

 

 

 

 

(from 20 Hz – 20 MHz)

 

 

 

 

CV peak-to-peak

4.5 mV

5 mV

6 mV

4.5 mV

CV rms

0.35 mV

1 mV

1 mV

0.35 mV

Load Effect (Regulation):

 

 

 

 

(for any output load change, with a maximum load-lead drop of 1V/lead)

 

 

Voltage

2 mV

2 mV

2 mV

0.5 mV

Current

2 mA

12 mA

5 mA

30 μA (@ 0 – 7 V)

 

 

 

 

65 μA (@ 7 – 50 V)

Source Effect (Regulation):

 

 

 

 

Voltage

1 mV

0.5 mV

1.2 mV

0.5 mV

Current

1 mA

5 mA

2 mA

30 μA

Programming Accuracy:

(@ 23 °C ±5 °C after 30 min. warm-up. Applies from min. to max. programming range)

Voltage high range

0.06% + 19 mV

0.06% + 10 mV

0.06% + 25 mV

0.016% + 6 mV

Voltage low range (5.5 V)

N/A

N/A

N/A

0.016% + 1.5 mV

Current high range

0.1% + 20 mA

0.1% + 30 mA

0.1% + 12 mA

0.04% + 200 μA

Current low range (100mA, @ 0 - 7 V)

N/A

N/A

N/A

0.04% + 15 μA

(100mA, @ 7 - 50 V)

N/A

N/A

N/A

0.04% + 55 μA

Measurement Accuracy: (at 23 °C ±5 °C)

Voltage high range

0.05% + 20 mV

0.05% + 10 mV

0.05% + 25 mV

0.016% + 6 mV

Voltage low range (5.5 V)

N/A

N/A

N/A

0.016% + 1.5 mV

Current high range

0.1% + 4 mA

0.1% + 30 mA

0.1% + 8 mA

0.04% + 160 μA

Current low range (100mA, @ 0 - 7 V)

N/A

N/A

N/A

0.03% + 15 μA NOTE 1

(100mA, @ 7 - 50 V)

N/A

N/A

N/A

0.03% + 55 μA

200 μA current range (Option 2UA)

N/A

N/A

N/A

0.5% + 100 nA

Load Transient Recovery Time: (time to recover to within the settling band following a load change

-from 60% to 100% and from 100% to 60% of full load for models N6751A & N6761A

-from 50% to 100% and from 100% to 50% of full load for models N6752A, N6762A, N6753A, & N6754A.)

Voltage settling band

± 75 mV NOTE 2

± 30 mV NOTE 3

± 90 mV NOTE 4

± 75 mV

Time

< 100 μs

< 100 μs

< 100 μs

< 100 μs

1Applies when measuring 4096 data points (SENSe:SWEep:POINts = 4096).

2When relay option 761 is installed, the settling band is ±125 mV for Model N6752A.

2When relay option 760 or 761 is installed, the settling band is ±200 mV for Model N6753A.

4When relay option 760 or 761 is installed, the settling band is ±350 mV for Model N6754A.

76

Series N6700 User’s Guide

Page 76
Image 76
Agilent Technologies N6700 manual Agilent Models N6751A/N6752A, N6753A/N6754A, N6761A/N6762A, Performance Specifications