Appendix C

 

Power Allocation

 

Power Limit Operation

96

Module Power Allocation

97

 

This chapter discusses the power allocation function.

 

For the majority of Agilent N6700 Modular Power System

 

configurations, full power is available from all installed power

 

modules. However, it is possible to configure a power system in

 

which the combined ratings of the power modules exceed the power

 

rating of the mainframe.

 

The power rating of Agilent N6700 mainframes is as follows:

 

Agilent N6700A/B = 400 W

 

Agilent N6701A = 600 W

 

Agilent N6702A @ nominal 100 - 120 VAC = 600 W

 

@ nominal 200 - 240 VAC = 1,200 W

 

Note that the power system will continue to operate normally as long

 

as the combined module output power is within the power rating of

 

the mainframe.

 

The power allocation function is only available on power system mainframes

NOTE

 

with firmware revision C.00.00 and up. To upgrade your power system with the

 

latest firmware revision, go to http://www.agilent.com/find/N6700firmware.

 

If this firmware is not installed, the later higher-power modules will not be

 

recognized by the mainframe and a self-test error will result.

 

 

Series N6700 User’s Guide

95

Page 95
Image 95
Agilent Technologies N6700 manual This chapter discusses the power allocation function, Rating of the mainframe, Mainframe