Agilent Technologies N6700 manual Module Power Allocation, To set a power limit on output

Models: N6700

1 109
Download 109 pages 3.15 Kb
Page 97
Image 97

Power Allocation Appendix C

Module Power Allocation

 

 

The following commands program the module power limit function:

 

 

 

 

 

 

Front Panel:

SCPI Command:

 

 

Select Output\Power.

To set a power limit on output 1:

 

 

Enter a power limit for each output.

POW:LIM 100,(@1)

 

 

 

 

 

To query the power limits that are set, send:

 

 

 

 

 

 

Front Panel:

SCPI Command:

 

 

Select Output\Power.

POW:LIM? (@1:4)

 

 

The power allocation for all output

 

 

 

channels is displayed in the dialog boxes.

 

 

 

To return all output channels to their default settings, you can either

 

 

cycle AC power or send the following commands:

 

 

 

 

 

 

Front Panel:

SCPI Command:

 

 

Select States\Reset

*RST

 

 

 

or

 

 

 

POW:LIM MAX,(@1:4)

 

 

 

 

 

When an output channel is set to MAX, it returns to its maximum rated value,

NOTE

 

 

and the power limit function will not activate.

 

 

 

 

Series N6700 User’s Guide