5Preventive Maintenance

Instrument Removal

NOTE

Detailed information related to rack mounting can be obtained in the

Agilent N8200A Series Synthetic Instrument Modules, Rack Configuration Guide,

 

 

N8200-90003.

 

 

 

To remove an instrument from the system, use one of the following procedures.

 

Required tools

 

#2 Phillips screwdriver

 

#2 POZIDRIV screwdriver

Standard instrument

To remove an instrument from a rack

Step

Notes

 

 

1 Turn off system power, but leave the system

If you do plan to turn computer power off

computer turned on.

for any reason, have the computer system

 

administrator:

 

Shut down all running software.

 

Shut down the computer.

 

 

2 Read “General Procedures and Techniques”,

Most cables are fairly easy to remove and

then disconnect all cables on the front and on

reconnect, and have no special

the rear panel.

considerations (besides making sure you

 

put the cables back in the right place).

 

Semirigid cables require more care,

 

especially when reconnecting them. Make

 

sure all semirigid cables, on the front and

 

back of an instrument are fully

 

disconnected before removing the unit.

 

 

3 When all cables are disconnected (including

 

the power cord), remove the screws in the

 

instrument’s rack “ears” that hold it in the

 

rack.

 

4Slide the instrument out.

If you feel any resistance when attempting to pull the instrument out, STOP! Look inside the cabinet and carefully examine all surrounding cables. Make sure all cables are fully disconnected.

98

Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz

Page 102
Image 102
Agilent Technologies N8201A manual Instrument Removal, Standard instrument, Required tools