Front and Rear Panel Features 4
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz 83
Line Power LED
The power indicator has the following states:
The green LED indicates when the downconverter standby switch is set to the on position.
The green LED is off when the switch is in the standby position.
The yellow LED indicates when the downconverter standby switch is set to the standby
position. The yellow LED is off when the switch is in the on position
LAN LED
The LAN LED indicator works in the following states:
State Power Status Illumination
OFF No power connected to rear p anel None
Standby Standby power. This is not a LINE
switch or power disconnecting device.
Solid amber
ON Power is on Solid green
State of the LAN LAN Status Illumination
ON Normal operation Solid green
ON Device identity. Blinking green
OFF LAN error fault:
No valid or duplicate IP address
Unable to obtain or renew previously
obtained DHCP lease.
Disconnected LAN cable
Solid Red
OFF This is the state w hen:
The system is initializing
or
A LAN reset has be initialized
None