Hardware Setup and Configuration

2

Measurement Procedure

4Tune the Spectrum Analyzer to the following:

Frequency: 7.5 MHz

Amplitude: 10 dBm

Span: 100 kHz

5Tune the downconverter to the following:

Center Frequency: 3.5 GHz

Attenuator: 20 dBm

RF Enabled: Enabled

Source: Internal

6Tune the Source to the following:

Frequency: 3.5 GHz

Amplitude: 0 dBm

RF On

Verify that a 7.5 MHz signal is present on the spectrum analyzer.

Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz

49

Page 53
Image 53
Agilent Technologies N8201A manual Measurement Procedure