Service, Support, and Safety Information

6

 

WA R N I N G

 

The mains wiring and connectors shall be compatible with the connector used in

 

 

the premise electrical system. Failure to ensure adequate earth grounding by not

 

 

 

 

 

 

 

 

using the correct components may cause product damage and serious injury.

 

 

 

Magnetic Susceptibility

 

 

 

 

Degradation of some product specifications can occur in the presence of ambient

 

CAUTION

 

 

 

power frequency magnetic fields of 30 A/m or greater. The product self-recovers

 

 

 

 

 

 

 

 

and operates as specified when removed or shielded from the ambient magnetic

 

 

 

 

fields.

 

 

 

 

Vibration

 

 

 

 

 

Degradation of some product specifications can occur if this instrument is operated

 

CAUTION

 

 

 

while subjected to continuous vibrations of less than or equal to 0.21 grms in the

 

 

 

 

 

 

 

 

5 Hz to 500 Hz frequency range.

 

 

 

Ground the Instrument or System

 

 

 

 

To minimize shock hazard, the instrument chassis and cover must be connected

 

WARNING

 

 

 

to an electrical protective earth ground. The instrument and/or system must be

 

 

 

 

 

 

 

 

connected to the AC power mains through a grounded power cable, with the

 

 

 

 

ground wire firmly connected to an electrical ground (safety ground) at the

 

 

 

 

power outlet. Any interruption of the protective (grounding) conductor or

 

 

 

 

disconnection of the protective earth terminal will cause a potential shock

 

 

 

 

hazard that could result in personal injury.

 

 

 

 

 

 

 

 

 

This is a Safety Class 1 Product (provided with a protective earthing ground

 

WARNING

 

 

 

 

incorporated in the power cord). The mains plug shall be inserted only in a socket

 

 

 

 

 

 

 

 

outlet provided with a protective earth contact. Any interruption of the protective

 

 

 

 

conductor inside or outside of the product is likely to make the product

 

 

 

 

dangerous. Intentional interruption is prohibited (IEC 348 clauses 17.3.3 c and

 

 

 

 

17.3.4).

 

 

 

 

 

 

 

 

 

Always use the three-prong AC power cord supplied with this product. Failure to

 

CAUTION

 

 

 

ensure adequate earth grounding by not using this cord may cause product damage.

 

 

 

 

 

 

 

 

 

Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz

107

Page 111
Image 111