Service, Support, and Safety Information 6
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz 107
Magnetic SusceptibilityVibrationGround the Instrument or System
WARNING
The mains wiring and connectors shall be compatible with the connector used in
the premise electrical system. Failure to ensure adequate earth grounding by not
using the correct components may cause product damage and serious injury.
CAUTION
Degradation of some product specifications can occur in the presence of ambient
power frequency magnetic fields of 30 A/m or greater. The product self-recovers
and operates as specified when removed or shielded from the ambient magneti c
fields.
CAUTION
Degradation of some product specifications can occur if this instrument is operated
while subjected to continuous vibrations of less than or equal to 0.21 grms in the
5 Hz to 500 Hz frequency range.
WARNING
To minimize shock hazard, the instrument chassis and cover must be connected
to an electrical protective earth ground. The instrument and/or system must be
connected to the AC power mains through a grounded power cable, with the
ground wire firmly connected to an electrical ground (safety ground) at the
power outlet. Any interruption of the protective (grounding) conductor or
disconnection of the protective earth terminal will cause a potential shock
hazard that could result in personal injury.
WARNING
This is a Safety Class 1 Product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall be inserted only in a socket
outlet provided with a protective earth contact. Any interruption of the protective
conductor inside or outside of the product is likely to make the product
dangerous. Intentional interruption is prohibited (IEC 348 clauses 17.3.3 c and
17.3.4).
CAUTION
Always use the three-prong AC power cord supplied with this product. Failure to
ensure adequate earth grounding by not using this cord may cause product damage.