46 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz
2Hardware Setup and Configuration
Measurement Procedure5Tune the Spectrum Analyzer to the following:
•Frequency: 7.5 MHz
•Amplitude: 10 dBm
•Span: 100 kHz
6Tune the downconverter to the following:
•Level: Attenuator: 20 dBm
•RF Enabled: Enabled
•Input RF: Signal Source: 50 MHz Cal
•Frequency: 50 MHz
7Verify that a 7.5 MHz signal is present on the spectrum analyzer.