Preventive Maintenance 5
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz 99
Half-Rack-Width InstrumentBench Top Instrument
To remove a half-width instrument from a system rack
1Power off the system. For details see the system installation
guide.
2 Remove the selected instrument’s power cord
from the power strip in the rack.
3The instrument is attached to the half-rack width
instrument beside it; remove that instrument’s
power cord from the power strip also.
The instruments are secured together
by lock links at the front and rear. The
lock links at the rear attach with screws.
The lock links at the front hook together.
4Remove the power cord and other cables from the
front and rear of both instruments.
Note the location of cables for
re-installation.
5Remove the four corner screws on the fron t of the
rack panel that secures the instruments in place.
The screws are located near the corners
of the face of the instrument.
Use a #2 Phillips screwdriver.
6Slide both instruments, as a single unit, out from
the front of the rack and set them on a secure, flat
surface.
To remove an instrument from a bench top system
1Power off each instrument in the system.
2 Unplug the selected instrument’s power
cord from the AC power supply.
3Remove the power cord and other cables
from the front and rear of the instrument.
Note the location of cables for re-installation.