2Hardware Setup and Configuration

If the Instrument was Unable to Join the LAN

or

If the LAN LED is Red

Possible Causes

Possible Solutions

The instrument is not connected to a LAN.

If connecting the instrument to a switch or hub, verify that the instrument is connected with a standard LAN cable.

An incorrect LAN cable is being used.

If connecting the instrument directly to a PC, verify that the instrument is connected with a cross-over cable.

If connecting the instrument to a switch or hub, verify that the instrument is connected with a standard LAN cable.

The device’s LAN port is not active.

Connect the instrument to a known working LAN port.

The device is configured to use DHCP, but no DHCP server is available.

Disable DHCP. Refer to “How to Set a Static IP Address" on page 56.

Connect the device to a LAN that uses a DHCP server.

The instrument is configured to use a duplicate static IP address.

Make sure that no other device is using the same IP address as your instrument.

Configure your instrument to use a different IP address. Refer to “How to Set a Static IP Address" on page 56.

62

Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz

Page 66
Image 66
Agilent Technologies N8201A manual Possible Causes