62 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz
2Hardware Setup and Configuration
If the Instrument was Unable to Join the LANor
If the LAN LED is RedSymptom
Possible Causes Possible Solutions
The instrument is not connected to a LAN. If connecting the instrument to a switch or hub, verify
that the instrument is connected with a standard LAN
cable.
An incorrect LAN cable is being used. •If connecting the instrument directly to a PC, verify
that the instrument is connected with a cross-over
cable.
•If connecting the instrument to a switch or hub,
verify that the instrument is connected with a
standard LAN cable.
The device’s LAN port is not active. Connect the instrument to a known working LAN
port.
The device is configured to use DHCP, but no DHCP
server is available.
•Disable DHCP. Refer to “How to Set a Static IP
Address" on page 56.
•Connect the device to a LAN that uses a DHCP
server.
The instrument is configured to use a duplicate static
IP address.
•Make sure that no other device is using the same
IP address as your instrument.
•Configure your instrument to use a different IP
address. Refer to “How to Set a Static IP
Address" on page 56.