Preventive Maintenance

5

Turn the silver nut clockwise by hand until it is snug, then tighten with an 8 inch-lb torque wrench (part number 8720-1765).

Bent Semirigid Cables

Semirigid cables are not intended to be bent outside of the factory. An accidental bend that is slight or gradual may be straightened carefully by hand. Semirigid cables that are crimped will affect system performance and must be replaced. Do not attempt to straighten a crimped semirigid cable.

Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz

97

Page 101
Image 101
Agilent Technologies N8201A manual Bent Semirigid Cables