106 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz
6Service, Support, and Safety Information
Environmental ConditionsBefore Applying Power
WARNING
DO NOT REMOVE AN INSTRUMENT COVER.
Operating personnel must not remove instrument covers. Component
replacement and internal adjustments must be made only by qualified service
personnel.
Instruments that appear damaged or defective should be made inoperative and
secured against unintended operation until they can be repaired by qualified
service personnel.
WARNING
If this product is not used as specified, the protection provided by the equipment
could be impaired. This product must be used only in a normal condition (in which
all means for protection are intact).
CAUTION
Unless otherwise noted in the specifications, this instrument or system is intended
for indoor use in an installation category II, pollution degree 2 environment per IEC
61010 second edition and 664 respectively. This instrument or system is designed
to operate at altitudes up to 4600 meters, and at temperatures between 0°C and
55 °C.
CAUTION
Ventilation Requirements: When installing the product in a cabinet, the convection
into and out of the product must not be restricted. The ambient temperature
(outside the cabinet) must be less than the maximum operating temperature of the
product by 4 °C for every 100 watts dissipated in the cabinet. If the total power
dissipated in the cabinet is greater than 800 watts, then forced convection must be
used.
CAUTION
Verify that the product is set to match the available line voltage, the correct-rating
service breaker is installed, the correct fuse is installed, and all safety precautions
are taken. Note the instrument externa l markings described in Table 6, “Safety symbols
and instrument markings,” on page 109.
Position the instrument so that it is not difficult to operate the disconnecting device.
This instrument has autoranging line voltage input, be sure the supply voltage is
within the specified range.