Banner L-GAGE manual GAGEQ45U Specifications, Response Curves, Output Protection Circuitry

Models: L-GAGE

1 66
Download 66 pages 29.94 Kb
Page 36
Image 36

LIGHT GAUGING

LABEL&

 

ULTRASONIC

 

SLOT

 

LIGHT

 

lUMINESCENCE

MEASURING

 

 

SCREENS

&

 

COLOR

 

TEMPERATURE

 

LIGHT-TO-

 

PICK

 

RADAR

 

MEASUREMENT & INSPECTION

U-GAGE®Q45U Specifications

Sensing Range

Near limit: 100 mm min.

Long Range: Near limit: 250 mm min.

 

 

Far limit: 1.4 m max.

 

Long Range: Far limit: 3.0 m max.

 

 

NOTE: The far limit may be extended on long range units, as far as 3.9 m for good acoustical targets

 

(hard surfaces with area greater than 100 cm2)

 

 

 

 

Supply Voltage and Current

Discrete: 12 to 24V dc (10% max. ripple); 100 mA (exclusive of load)

 

 

Analog: 15 to 24V dc (10% max. ripple); 100 mA (exclusive of load)

 

 

 

 

 

 

Ultrasonic Frequency

Long Range: 128 kHz

 

Short Range: 230 kHz

 

 

 

 

Supply Protection Circuitry

Protected against reverse polarity and transient voltages.

 

 

 

Output Protection Circuitry

Protected against false pulse on power-up and continuous overload or short-circuit of outputs.

 

 

Output Configuration

Discrete: Bipolar: One current sourcing (PNP) and one current sinking (NPN) open-collector transistor.

 

Analog: One voltage sourcing and one current sourcing; one or the other output is enabled by internal

 

programming switch #2.

 

 

 

 

 

 

Output Ratings

Discrete: 150 mA max. (each)

 

 

 

OFF-state leakage current: less than 25 µA at 24V dc

 

 

ON-state saturation voltage: less than 1.5V at 10 mA; less than 2.0V at 150 mA

 

Analog: Voltage sourcing: 0 to 10V dc, 10 mA max.

 

 

Current sourcing: 4 to 20 mA, 1 to 500 Ω impedance

 

 

 

 

 

Performance Specifications

Analog resolution or

Short Range

Long Range

 

 

 

 

 

discrete repeatability: ± 0.1% of sensing distance

± 0.1% of sensing distance

 

Analog Linearity:

(± 0.25 mm min.)

(± 0.5 mm min.)

 

1% of full scale

1% of full scale

 

Temperature effect: 0.05% of sensing distance/ ° C with temp. comp.

0.05% of sensing distance/° C

 

Min. window size:

0.2% of sensing distance/ ° C without temp. comp.

 

 

10 mm

 

25 mm

 

Hysteresis (discrete output): 5 mm

10 mm

 

 

 

Response Curves

Short Range: See charts RC-2 and RC-4 on page 516.

 

 

Long Range: See charts RC-3 and RC-5 on page 516.

 

Adjustments

The following may be selected by a 4-position DIP switch located on top of the sensor, beneath a

 

transparent o-ring sealed acrylic cover:

 

 

Discrete: Switch 1: Output normally open/normally closed (pump in/pump out)

 

Switch 2: High/Low level control mode or on/off presence sensing mode

 

Switch 3 & 4: Response speed selection (digital filter)

 

 

Analog: Switch 1: Output slope positive or output slope negative

 

 

Switch 2: Current output mode or voltage output mode

 

 

Switch 3: Loss of echo min/max mode or loss of echo Hold Mode

 

Switch 4: Loss of echo min/max default output value

 

 

 

 

Indicators

Discrete: Three status LEDs:

 

 

Green ON steady: power to sensor is ON

 

 

Green flashing: output is overloaded

 

 

Yellow ON steady: outputs are conducting (Yellow LED also indicates programming

 

 

 

status during setup mode)

 

 

Red flashing: indicates relative strength of received echo

 

 

Analog: Three status LEDs:

 

 

 

Green ON steady: power to sensor is ON

 

 

Green flashing: current output fault detected (the 4-20 mA current path to ground has

 

 

been opened)

 

 

Yellow ON steady: target is sensed within the window limits (Yellow LED also indicates

 

 

 

programming status during setup mode)

 

 

Red flashing: indicates relative strength of received echo

 

 

5-segment moving dot LED indicates the position of the target within the sensing window.

 

 

 

 

 

 

 

 

 

More on

 

 

 

 

next page

278 More information online at bannerengineering.com

Page 36
Image 36
Banner L-GAGE manual GAGEQ45U Specifications, Response Curves, Output Protection Circuitry, Performance Specifications