Conair MPW, MPA manual Performing An Auto Tune, b a d, Insufficient setpoint/process deviation

Models: MPW MPA

1 86
Download 86 pages 25.93 Kb
Page 41
Image 41
Insufficient setpoint/process deviation.

PERFORMING AN AUTO TUNE

5

Press Run and Invalid PID constants were generated. (the Setpoint Select button) simultaneously to begin the Auto Tune.

The Actual display will flash “At” and the current process temperature to indicate that an Auto Tune is underway.

If Auto Tune is successful, the controller automatically starts controlling using the new PID parameters.

If you press the STOP button or a fault occurs during the Auto Tune, the control enters stop mode and Auto Tuning immediately terminates. The actual display stops flashing “At.” If a fault occurred, the control will display the appro- priate error message.

ERROR MESSAGE

 

ERROR DESCRIPTION

 

 

 

 

 

 

 

 

 

Insufficient setpoint/process deviation.

 

 

 

 

 

 

 

 

 

 

A t

 

 

 

d E v

 

 

 

 

 

 

 

 

If the absolute value of (setpoint - process

 

 

 

 

 

 

 

 

 

value) is less than 25° F. The Auto Tune can-

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

not be started until the temperature

 

 

 

 

 

 

 

 

 

difference is at least 25° F.

 

 

 

 

 

 

 

 

 

Auto Tune timed out. Auto Tune will time out

 

 

 

 

 

 

 

 

 

 

 

A t

 

 

 

t I

 

 

 

 

 

 

 

 

 

if a stable process value cannot be obtained

 

 

 

 

 

 

 

 

 

5 minutes into the tune, or if the tuning

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

process takes longer than 30 minutes. If this

 

 

 

 

 

 

 

 

 

error occurs, verify that you followed every

 

 

 

 

 

 

 

 

 

requirement under “To ensure a successful

 

 

 

 

 

 

 

 

 

Auto Tune” and perform a second tune.

 

 

 

 

 

 

 

 

 

Invalid PID constants were generated.

 

 

A t

 

 

 

b a d

 

 

 

 

 

 

 

 

 

The most likely causes of this error is a tune

 

 

 

 

 

 

 

 

 

 

started inappropriately or an external ele- ment (i.e., loose thermocouple) that upset the process while tuning was in progress. Verify that you followed every requirement under “To ensure a successful Auto Tune” and perform a second tune.

UGH004/0999

microKool Portable Chillers

OPERATION 4-9