World Class 3000

Table A-1. Specifications for Oxygen Analyzing Equipment.1, 2

Probe lengths, nominal

18 inches (457 mm), 3 feet (0.91 m), 6 feet

 

(1.83 m), 9 feet (2.74 m), or 12 feet (3.66 m),

Temperature limits in process

depending on duct dimensions

 

measurement area

50° to 1300°F (10° to 704°C)

Standard/current output

4-20 mA dc signal (factory set)

O2 indication (Digital display

 

and analog output)

0.1% O2 or ±3% of reading, whichever is

 

greater using Rosemount Analytical calibration

 

gases

System speed of response

less than 3 seconds (amplifier output)

Resolution sensitivity

0.01% O2 transmitted signal

HPS 3000 housing

NEMA 4X (IP56)

Probe reference air flow

2 scfh (56.6 L/hr) clean, dry, instrument quality

Calibration gas mixtures

air (20.95% O2), regulated to 5 psi (34 kPa)

Rosemount Hagan Calibration Gas Kit Part No.

 

6296A27G01 contains 0.4% O2N2 Nominal and

Calibration gas flow

8% O2N2 Nominal

5 scfh (141.6 L/hr)

HPS 3000 Power supply

100/110/220 ±10% Vac at 50/60 Hz

HPS 3000 Power requirement

200 VA

HPS 3000 Ambient Operating Temperature

32° to 120°F (0° to 50°C)

Ambient operating temperature (Probe Junction Box)

300°F (150°C) max

Approximate shipping weights:

 

18 inch (457 mm) package

55 pounds (24.97 kg)

3 foot (0.91 m) package

60 pounds (27.24 kg)

6 foot (1.83 m) package

65 pounds (29.51 kg)

9 foot (2.74 m) package

72 pounds (32.66 kg)

12 foot (3.66 m) package

78 pounds (35.38 kg)

1All static performance characteristics are with operating variables constant.

2Equipment ordered utilizing this document as reference will be supplied to the USA standard design. Custom- ers requiring the EEC standard design should request the EEC documentation and utilize its ordering data. Temperatures over 1000°F (537°C) may affect the ease of field cell replaceability.

A-2 Appendices

Rosemount Analytical Inc. A Division of Emerson Process Management