Emerson 3000 World Class, Instruction Manual, c.Reference Air Package, d.Service Required

Models: 3000

1 184
Download 184 pages 52.9 Kb
Page 44
Image 44
c.Reference Air Package

Instruction Manual

IB-106-300NH Rev. 4.3 May 2005

World Class 3000

c.Reference Air Package

After the oxygen analyzing (probe) unit is installed, connect the reference air set to the probe junction box. The reference air set should be installed in accordance with Figure 2-3.

d.Service Required

1.Power input: 100, 115 or 220 Vac sin- gle phase, 50 to 60 Hz, 3 amp mini- mum. (See label.)

2.Compressed air: 10 psig (68.95 kPa) minimum, 225 psig (1551.38 kPa)

maximum at 2 scfh (56.6 L/hr) maxi- mum; supplied by one of the following (less than 40 parts-per-million total hy- drocarbons). Regulator outlet pressure should be set at 5 psi (35 kPa).

(a)Instrument air - clean, dry.

(b)Bottled standard air with step-down regulator.

(c)Bottled compressed gas mixture (20.95% oxygen in nitrogen).

(d)Other equivalent clean, dry, oil-free air supply.

 

0.125-27 NPT FEMALE

 

 

3.12 (79.25) MAX

NOTE: DIMENSIONS ARE IN INCHES WITH

 

OUTLET CONNECTION

 

 

 

 

 

 

 

2.250 (57.15)

 

MILLIMETERS IN PARENTHESES.

 

1

 

 

3

 

 

 

2

OUTLET

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4.81

(122.17)

 

 

 

 

0.25-18 NPT FEMALE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

INLET CONNECTION

8.50 MAX

 

 

 

 

 

 

 

 

 

FLOW SET

 

 

 

 

 

 

 

 

(215.90)

 

 

 

 

 

 

 

 

 

 

POINT KNOB

 

1.19

 

 

 

2.0

 

 

 

 

 

 

 

 

 

2 MOUNTING HOLES

 

 

 

(30.22)

 

 

 

 

 

 

 

(50.80)

 

 

 

 

 

 

 

 

3.19 (81.03)

LG

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.50

THROUGH BODY FOR

 

 

 

DRAIN VALVE

 

(38.10)

0.312 (7.92) DIA BOLTS

 

 

 

 

 

 

 

 

 

 

 

 

10.0 REF

 

 

 

 

 

 

 

 

 

 

(254)

 

 

 

 

 

 

 

0.250 OR 6 MM OD

 

 

 

 

 

 

 

 

TUBE COMPRESSION

 

 

1

FLOWMETER

 

 

0.2-2.0 SCFH

771B635H02

FITTING (SUPPLIED BY WECO)

 

2

2" PRESSURE GAGE

0-15 PSIG

275431-006

 

 

TO PROBE HEAD

 

 

3

COMBINATION FILTER-REG.

0-30 PSIG

4505C21G01

 

 

 

 

 

0.250 OR 6 MM OD TUBING (SUPPLIED BY CUSTOMER)

REF AIR SET

263C152G01

COMPRESSED AIR SUPPLY 10-225 PSIG MAX PRESSURE

SCHEMATIC HOOKUP FOR REFERENCE AIR SUPPLY ON OXYGEN ANALYZER PROBE HEAD.

27270003

Figure 2-3. Air Set, Plant Air Connection

2-8 Installation

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 44
Image 44
Emerson 3000 World Class, Instruction Manual, c.Reference Air Package, d.Service Required, 3.Air Set, Plant Air Connection