SYMPTOM - MICROPROCESSOR BOARD LED IS STEADY ON

Instruction Bulletin

Appendix E Rev. 4.6 May 2005

World Class 3000

SYMPTOM - MICROPROCESSOR BOARD LED IS STEADY ON

OXYGEN ANALYZER

YES

SYSTEM IS EQUIPPED

 

 

WITH HPS 3000.

 

 

 

 

 

NO

 

 

 

 

CHECK JUMPERS JM6 ON

 

 

MICROPROCESSOR BOARD

 

AND JM1 ON INTERCONNECT

NO

BOARD ARE INSTALLED.

 

 

CHECK JUMPERS 9

 

OR 10 ON POWER

 

SUPPLY BOARD FOR COR-

 

RECT CONFIGURATIONS.

 

 

 

 

 

 

YES

 

 

 

 

 

REFER TO PROBE

TROUBLESHOOTING,

APPENDIX A, SECTION 2.

CHECK THAT JUMPERS

JM6 ON MICROPROCESSOR BOARD AND JM1 ON INTERCONNECT BOARD ARE REMOVED.

DISCONNECT POWER AND INSTALL JUMPERS CORRECTLY. REFER TO FIGURE 2-5 OF MAIN IB SECTION.

REFER TO APPENDIX B, SECTION 2, FOR JUMPER LOCATIONS.

27270021

Figure E-4. IFT Troubleshooting Flowchart, #2

END OF FLOWCHART #2

SYMPTOM - GENERAL USER INTERFACE OR

LED DISPLAY PANEL DOES NOT FUNCTION

REPLACE GUI BOARD.

STILL NOT FUNCTIONING.

CALL FACTORY FOR

ASSISTANCE.