2-4 MULTIPROBE CALIBRATION GAS

World Class 3000

Instruction Manual

IB-106-300NH Rev. 4.3 May 2005

2-4 MULTIPROBE CALIBRATION GAS

SEQUENCER INSTALLATION

a.Mechanical Installation

The outline drawing of the MPS module in Figure 2-17 shows mounting centers and clearances. The box is designed to be mounted on a wall or bulkhead. The MPS module should be installed no further than 300 feet (91 mz) piping distance from the probe, and no more than 1000 feet (303 m) cabling distance from the IFT. Install the MPS module in a location where the ambi- ent temperature is between -20° and 160°F (-30° and 71°C).

b.Gas Connections

Figure 2-18 shows the bottom of the MPS where the gas connections are made. 1/4 inch threaded fittings are used.

1.Connect the reference air supply to INSTR. AIR IN. The air pressure regu- lator valve is set at the factory to 20 psi (138 kPa). If the reference air pressure should need readjustment, turn the knob on the top of the valve until the desired pressure is obtained.

2.Connect the high O2 calibration gas to HIGH GAS. The calibration gas pres- sure should be set at 20 psi (138 kPa).

12.00

(304.80)

10.00

(254.00)

Analytical

SEQUENCER INSTALLATION 14.00 (355.60) REF b.Gas Connections

 

 

PROBE 1

PROBE 2

PROBE 3

PROBE4

HIGH CAL

LOW CAL

CAL GAS

CAL GAS

CAL GAS

CAL GAS

GAS IN

GAS IN

OUT

OUT

OUT

OUT

 

 

REF AIR

REF AIR

REF AIR

REF AIR

 

 

OUT

OUT

OUT

OUT

INSTR

 

 

 

 

 

AIR

 

 

 

 

 

12.00

(304.80)

NOTE: DIMENSIONS ARE IN INCHES

WITH MILLIMETERS IN PARENTHESES.

Analytical 0.84 (21.34)

Figure 2-17. MPS ModuleManual background 1.96 (49.78)

Manual backgroundManual background 3.09 (78.49)

Manual backgroundManual background 4.21 (106.93)

Manual background 5.25 (133.35)

5.54 (140.72)

37840010

Figure 2-17. MPS Module

Rosemount Analytical Inc. A Division of Emerson Process Management