SECTION 4 CALIBRATION

4-1 ANALOG OUTPUT CALIBRATION

For the analog output to perform within the pub- lished specifications, it must be manually cali- brated. The only equipment needed to perform the calibration is a voltage or current meter, de- pending on which mode of operation is to be calibrated. Prior to manual calibration, remove the IFT from any control loops it may be in.

Prior to manual calibration, the IFT should be removed from any auto- matic control loops. Failure to remove the IFT from control loops prior to calibration may result in faulty equip- ment performance.

Once initiated from the Setup - Analog Outputs menu, the calibration procedure is self guiding.

4-2 SYSTEM CALIBRATION

a.Overview

The primary purpose of an oxygen analyzer is to give an accurate representation of the percentage of O2 in the gas stream. The system should be calibrated periodically to maintain an accuracy which may otherwise be reduced over time due to cell aging. A calibration record sheet is provided at the end of this section to track cell performance.

A requirement for calibration is a set of two accurate calibration gases spanning the oxygen range of most interest. For example, 0.4% and 8% for a 0-10% oxygen range.

Under normal conditions the probe should not need frequent calibration. Because cali- bration is necessary, the system can be equipped with the optional MPS 3000 Multi- probe Calibration Gas Sequencer for fully automatic calibration at regular intervals. Without an MPS, the probes must be cali- brated manually (semiautomatically).

b.Probe Calibration

1.Previous Calibration Constants Functionality

There are three sets of registers used to store calibration constants. These are: Latest Calibration, Previous Cali- bration, and Calculation. Only the val- ues in the Calculation register are used to calculate the oxygen value for dis- play and representation on the analog output signal. These values may be changed in two ways.

(a)The operator may change the val- ues through the SETUP menu. The operator may adjust the slope and constant individually, or reset both to the values calculated during the last good calibration. To reset the values, move the cursor to RESET SLOPE & CONST and push

ENTER.

(b)The IFT will automatically change the values after each calibration as follows:

The values in the Latest Calibra- tion registers are updated after every complete calibration, even if the calibration is not successful. If the calibration is successful, the values in the Latest Calibration registers are copied into the Previ- ous Calibration registers. This is accomplished prior to the update of the Latest Calibration registers. The new slope and constant are copied into the Calculation register.

If the calibration fails, the Previous Calibration registers retain their existing values, while the Latest Calibration registers record the values of the failed calibration. The Calculation register is not updated when the calibration fails.

Rosemount Analytical Inc. A Division of Emerson Process Management

Calibration 4-1

Page 63
Image 63
Emerson 3000 instruction manual Analog Output Calibration, System Calibration, Overview

3000 specifications

The Emerson 3000 is a cutting-edge control system designed to enhance the efficiency, reliability, and precision of industrial operations. Employed across various sectors such as oil and gas, pharmaceutical, food and beverage, and power generation, the Emerson 3000 has gained recognition for its robustness and versatility.

One of the main features of the Emerson 3000 is its advanced process control capability. With integrated control algorithms, it can optimize complex processes in real-time, resulting in significant improvements in production rates and reduced operational costs. The system's predictive analytics capabilities enable operators to anticipate equipment failures and maintenance needs, allowing for proactive management and minimizing downtime.

The Emerson 3000 features a modular architecture, providing flexibility for scaling and customization. Operators can easily tailor the system to fit specific application needs, whether it requires additional control loops or integration with other systems. This adaptability is particularly beneficial for facilities planning for future expansions or modifications.

Another technology highlight of the Emerson 3000 is its seamless integration with the latest Internet of Things (IoT) advancements. The system is designed to communicate effectively with a variety of smart devices and sensors, harnessing data to create insightful analytics that drive operational excellence. This connectivity empowers businesses to leverage big data for improved decision-making and increased agility.

Additionally, the Emerson 3000 incorporates state-of-the-art cybersecurity measures to safeguard critical data and operations. With built-in security protocols and regular updates, the system protects against emerging cyber threats, ensuring the integrity of the control network.

User experience is also a focal point of the Emerson 3000. The intuitive graphical user interface presents complex data in a user-friendly format, making it easier for operators to monitor system performance and respond to alerts quickly. This ease of use contributes to enhanced safety and operational efficiency.

In summary, the Emerson 3000 represents a fusion of advanced process control, modular design, IoT connectivity, robust cybersecurity, and user-centric interface, making it an ideal choice for industries seeking to enhance their operational performance while adapting to ever-evolving technological landscapes.