Emerson 3000 Safety Instructions, For The Wiring And Installation Of This Apparatus, World Class

Models: 3000

1 184
Download 184 pages 52.9 Kb
Page 20
Image 20
SAFETY INSTRUCTIONS

Instruction Manual

IB-106-300NH Rev. 4.3 May 2005

World Class 3000

IMPORTANT

SAFETY INSTRUCTIONS

FOR THE WIRING AND INSTALLATION

OF THIS APPARATUS

The following safety instructions apply specifically to all EU member states. They should be strictly adhered to in order to assure compliance with the Low Voltage Directive. Non- EU states should also comply with the following unless superseded by local or National Standards.

1.Adequate earth connections should be made to all earthing points, internal and external, where provided.

2.After installation or troubleshooting, all safety covers and safety grounds must be replaced. The integrity of all earth terminals must be maintained at all times.

3.Mains supply cords should comply with the requirements of IEC227 or IEC245.

4.All wiring shall be suitable for use in an ambient temperature of greater than 75°C.

5.All cable glands used should be of such internal dimensions as to provide adequate cable anchorage.

6.To ensure safe operation of this equipment, connection to the mains supply should only be made through a circuit breaker which will disconnect all circuits carrying conductors during a fault situation. The circuit breaker may also include a mechanically operated isolating switch. If not, then another means of disconnecting the equipment from the supply must be provided and clearly marked as such. Circuit breakers or switches must comply with a recognized standard such as IEC947. All wiring must conform with any local standards.

7.Where equipment or covers are marked with the symbol to the right, hazard- ous voltages are likely to be present beneath. These covers should only be removed when power is removed from the equipment — and then only by trained service personnel.

8.Where equipment or covers are marked with the symbol to the right, there is a danger from hot surfaces beneath. These covers should only be removed by trained service personnel when power is removed from the equipment. Cer- tain surfaces may remain hot to the touch.

9.Where equipment or covers are marked with the symbol to the right, refer to the Operator Manual for instructions.

10.All graphical symbols used in this product are from one or more of the follow- ing standards: EN61010-1, IEC417, and ISO3864.

P-2

Rosemount Analytical Inc. A Division of Emerson Process Management