Instruction Manual

IB-106-300NH Rev. 4.3 May 2005

World Class 3000

Do not install jumper JM6 on the mi- croprocessor board, or JM1 on the in- terconnect board, if an HPS is installed in the system. This will result in system failure.

3.The IFT must have JM6 on the micro- processor board (Figure 2-8 and Figure 2-9) and JM1 on the interconnect board (Figure 2-10 and Figure 2-11) installed if an HPS is not installed in the system.

4.If an MPS is not used in the system, wire jumper between CAL RET and NO GAS must be installed on the in- terconnect board. Remove wire jumper

if MPS is installed in the system. Refer to Figure 2-7, note 6.

5.The power cable should comply with the safety regulations in the user's country and should not be smaller than 16 gauge, 3 amp.

6.Before supplying power to the IFT, ver- ify that the jumpers have been properly set in the IFT (Figure 2-5, Figure 2-8, and Figure 2-10).

7.Terminal strip J5 on the power supply board is used for supplying the IFT with power. Terminal strip J6 on the power supply board is used to supply the probe heater with power if an HPS is not used (Figure 2-6).

 

 

 

 

 

 

 

 

ALWAYS DISCONNECT LINE VOLTAGE

 

 

 

 

 

 

 

 

 

 

 

JUMPER

 

 

FROM INTELLIGENT FIELD TRANSMITTER

 

 

 

CONFIGURATION

 

BEFORE CHANGING JUMPERS.

 

 

 

 

 

 

 

 

 

 

LINE VOLTAGE

JUMPER

 

 

 

 

 

PROBE HEATER

JUMPER

SELECTION

(INSTALL)

 

 

 

 

VOLTAGE SELECTION

(INSTALL)

 

 

 

 

 

 

 

 

 

100 V.A.C.

JM3, JM7, JM2

 

 

 

WORLD CLASS PROBE (44V)

JM10

 

 

 

 

 

 

 

 

 

 

120 V.A.C.

JM8, JM7, JM1

 

 

 

 

218 PROBE (115V)

JM9

 

 

 

 

 

 

 

 

 

 

220 V.A.C.

JM6, JM5, JM2

 

 

 

WORLD CLASS "DIRECT

JM9

 

 

 

 

 

 

240 V.A.C.

JM6, JM5, JM1

 

 

REPLACEMENT" PROBE (115V)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

21190012

If incorrect heater voltage is selected, damage to the probe may occur. For HPS voltage se- lection jumper, refer to Figure 2-15. Always update the relevant labeling to reflect the set voltage.

Figure 2-5. Power Supply Board Jumper Configuration