Emerson 3000 instruction manual World Class, Instruction Manual, e. Fully Automatic Calibration

Models: 3000

1 184
Download 184 pages 52.9 Kb
Page 68
Image 68
e. Fully Automatic Calibration

Instruction Manual

IB-106-300NH Rev. 4.3

May 2005

 

 

 

 

 

 

 

 

 

REFERENCE AIR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CALIBRATE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CONNECTION

IN PLACE

 

 

 

 

 

 

 

 

 

 

 

 

 

CONNECTION

CHECK

CALIBRATION

 

VALVE

GAS HOSE

 

 

CONNECTS

 

 

TO CHECK

 

 

VALVE

World Class 3000

ENTER to start. Follow the data queues. Refer to Table 5-4, CALI- BRATE O2 Sub-menu.

e. Fully Automatic Calibration

1. Calibration Gases For Fully Automatic

Calibration. For fully automatic calibra-

PUSHBUTTON

 

REGULATOR

2

WITH CONTENTS

O

GAGE - SET 5 SCFH

0.4%

 

NOTE: PROBE CALIBRATION GAS FITTING HAS A SEAL CAP WHICH MUST BE IN PLACE EXCEPT DURING CALIBRATION.

2 O8.0%

37840013

tion, an MPS 3000 Multiprobe Calibra-

tion Gas Sequencer is required as well

as the two types of calibration gas.

Do not use 100% nitrogen as a zero gas. It is suggested that gas for the

Figure 4-3. Typical Portable Calibration Setup

d.Manual (Semiautomatic) Calibration Procedure

The following procedure relates to an op- erator initiated calibration selected at the IFT by pressing the CAL key. The calibra- tion is manually performed by the operator upon data queues from the IFT. Any system without an MPS 3000 Multiprobe Calibration Gas Sequencer must follow these steps.

1.Press SETUP to display the SETUP menu. Select PROBE CALIBRATION sub-menu. Ensure that Auto Cal is dis- abled. Set the cursor on Auto Cal. Press ENTER. Set Auto Cal to NO if not already done.

2.Press the CAL key. Select PERFORM CALIBRATION sub-menu. "Press EN- TER to start Manual Calibration" will appear on the LCD display. Press

zero be between 0.4% and 2.0% O2. Do not use gases with hydrocarbon con- centrations of more than 40 parts per million. Failure to use proper gases will result in erroneous readings.

NOTE

Ambient air is not recommended for use as high calibration gas. An 8% O2 balance in nitrogen is recommended for high calibration gas.

Two tanks of precision calibration gas mixtures. Recommended calibration gases are nominally 0.4% and 8.0% oxygen in nitrogen set calibration gas pressure at 20 psi (138 kPa).

A typical automatic calibration system is shown in Figure 4-4.

4-6 Calibration

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 68
Image 68
Emerson 3000 World Class, Instruction Manual, e. Fully Automatic Calibration, 3.Typical Portable Calibration Setup