Instruction Manual

IB-106-300NH Rev. 4.3 May 2005

World Class 3000

2. Calibration Methods

Test Method "A" Fixed Tanks and

 

Manifolds

There are three calibration methods: manual (semiautomatic), manually ini- tiated automatic, and fully automatic. Manual (semiautomatic) calibration is done without an MPS unit. Calibration gases are switched on and off by the operator and the IFT is sequenced through the calibration procedure by the operator with the front panel key- board. The IFT prompts the operator for the correct action. Manually initiated automatic calibration is done with an MPS. The operator manually initiates the calibration at the IFT or through a remote switch, and the IFT controls the operation of the MPS unit and the cali- bration sequencing. Fully automatic calibration requires no action from the operator. The setup is the same as semiautomatic except the IFT auto- matically initiates the calibration at a fixed calibration interval. In this mode the operator can also manually initiate calibrations between the intervals in the same manner as semiautomatic calibrations.

c.Manual (Semiautomatic) Calibration

1.Calibration Gases For Manual (Semiautomatic) Calibration

There are two options for supplying calibration gases to the probe during semiautomatic calibration. The first "A" uses refillable bottles and adjustable 2- stage pressure regulators; the second "B" uses disposable bottles and a fixed single stage regulator to provide a mixed flow. Normally, the first (method "A") will have a higher cost and not be portable. The second ("B") is less costly, portable, and weighs about 10 lbs (4.5 kg).

(a) Required Equipment

Do not use 100% nitrogen as a zero gas. It is suggested that gas for the zero be between 0.4% and 2.0% O2. Do not use gases with hydrocarbon con- centrations of more than 40 parts per million. Failure to use proper gases will result in erroneous readings.

NOTE

Ambient air is not recommended for use as high calibration gas. An 8% O2 balance in nitrogen is recommended for high calibration gas.

1Two tanks of precision cali- bration gas mixtures. Rec- ommended calibration gases are nominally 0.4% and 8.0% oxygen in nitrogen.

Two sources of calibrated gas mixtures are:

LIQUID CARBONIC GAS CORP.

SPECIALTY GAS LABORATORIES

700 South Alameda Street Los Angeles, California 90058 213/585-2154

767 Industrial Road

San Carlos, California 94070 415/592-7303

9950 Chemical Road Pasadena, Texas 77507 713/474-4141

12054 S.W. Doty Avenue Chicago, Illinois 60628 312/568-8840

603 Bergen Street

Harrison, New Jersey 07029 201/485-1995

4-2 Calibration

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 64
Image 64
Emerson 3000 instruction manual Calibration Methods, Liquid Carbonic GAS Corp Specialty GAS Laboratories

3000 specifications

The Emerson 3000 is a cutting-edge control system designed to enhance the efficiency, reliability, and precision of industrial operations. Employed across various sectors such as oil and gas, pharmaceutical, food and beverage, and power generation, the Emerson 3000 has gained recognition for its robustness and versatility.

One of the main features of the Emerson 3000 is its advanced process control capability. With integrated control algorithms, it can optimize complex processes in real-time, resulting in significant improvements in production rates and reduced operational costs. The system's predictive analytics capabilities enable operators to anticipate equipment failures and maintenance needs, allowing for proactive management and minimizing downtime.

The Emerson 3000 features a modular architecture, providing flexibility for scaling and customization. Operators can easily tailor the system to fit specific application needs, whether it requires additional control loops or integration with other systems. This adaptability is particularly beneficial for facilities planning for future expansions or modifications.

Another technology highlight of the Emerson 3000 is its seamless integration with the latest Internet of Things (IoT) advancements. The system is designed to communicate effectively with a variety of smart devices and sensors, harnessing data to create insightful analytics that drive operational excellence. This connectivity empowers businesses to leverage big data for improved decision-making and increased agility.

Additionally, the Emerson 3000 incorporates state-of-the-art cybersecurity measures to safeguard critical data and operations. With built-in security protocols and regular updates, the system protects against emerging cyber threats, ensuring the integrity of the control network.

User experience is also a focal point of the Emerson 3000. The intuitive graphical user interface presents complex data in a user-friendly format, making it easier for operators to monitor system performance and respond to alerts quickly. This ease of use contributes to enhanced safety and operational efficiency.

In summary, the Emerson 3000 represents a fusion of advanced process control, modular design, IoT connectivity, robust cybersecurity, and user-centric interface, making it an ideal choice for industries seeking to enhance their operational performance while adapting to ever-evolving technological landscapes.